TSN 910 SC
Dimension and weight :
- Appr. 99 W x 59 D x 137 H cm, ± 60 Kg (cabinet type)
- Appr. 99 W x 59 D x 137 H cm, ± 50 Kg (trolley type)
Guard and handle type Steel
Castors 4 inches, two lockable
IV pole Adjustable with fixation clamp
Tray for instrument with max load ± 2,5 Kg
Dimension Appr. 86 W x 41 D x 45 H cm
Mattress size Appr. 65 cm x 35 cm
Mattress to hood Appr. 36 cm
Mattress elevation 0 to12° (trendelenburg and reverse by crank)
Mattress tray ABS
Access type 4 port doors with cuff & ABS latches, 2 iris port, 4 tubing grommets, 1 access panel
Control panel Hidden keys type, auto self-test system, LED seven-segment
Control mode Manual (Air mode only)
Controller type Microcontroller
Control panel lock Automatic key lock after 10 seconds unchanged
Temperature display range 17.1 °C – 42.4 °C
Air mode control range 25.0 °C – 37.0 °C, resolution 0.1 °C
Air mode maximum override 39.0 °C
Warm up time ≤ 30 minutes
Humidity range 15 – 90 % RH
Water reservoir capacity 1500 ml
Noise level in hood ≤ 60 dB
Oxygen supply Oxygen inlet with micro filter
Air flow velocity ≤ 10 cm/s
Alarm indicator visible (dot matrix LCD) and audible (buzzer)
Alarm silencer 2 minutes for Low Internal Battery or Mains Failure, 10 minutes for others
Information on LCD Heater level, alarms, serial numbers, internal battery level and User ID (by customer request)
• System failure
• Airflow failure
• Low internal battery
• High temperature (38.0° C or 40.0° C)
• Sensor failure
• Air over temperature (37.5° C or 39.5° C)
• Mains failure
• Warm-up failure (> 30 minutes)
• Blower failure
• Air deviation ±3°C
• Portable phototherapy ( TLC 2 HP)
• Oxyhood (TSN 046 series)
• Digital baby scale ( TSN 9806 DBS)
• X-ray cassette tray
• Double wall