TSN 910 SC EHL
Power AC 220 V, 50 Hz, 350 Watt
Dimension Appr. 110 W x 59 D x (130 – 150) H cm
Weight Appr.105 Kg (default) please refer to optional stands
Guard and handle type PU and Steel
Castors 4 inches, two lockable
IV pole Adjustable with fixation clamp, stainless steel
Tray for instrument with max load ± 2,5 Kg
Dimension Appr. 86 W x 41 D x 45 H cm
Mattress size Appr. 65 cm x 35 cm
Mattress to hood Appr. 36 cm
Mattress elevation 0 to12° (trendelenburg and reverse by crank)
Mattress tray ABS
Access type 4 port doors with cuff & ABS latches, 2 iris port, 4 tubing grommets, 1 access panel
Control panel Membran Keyboard type, auto self-test system, arge LED seven-segment for actual temperature
Control mode Servo (skin Mode) and manual (Air mode only)
Controller type Microcontroller
Control panel lock Automatic key lock after 10 seconds unchanged
Temperature display range 17.1 °C – 42.4 °C
Air mode control range 25.0 °C – 37.0 °C, resolution 0.1 °C
Air mode maximum override 39.0 °C
Warm up time ≤ 30 minutes
Humidity range 15 – 90 % RH
Water reservoir capacity 1500 ml
Noise level in hood ≤ 60 dB
Oxygen supply Oxygen inlet with micro filter
Air flow velocity ≤ 10 cm/s
Alarm indicator visible (dot matrix LCD) and audible (buzzer)
Alarm silencer 2 minutes for Low Internal Battery or Mains Failure, 10 minutes for others
Information on LCD Heater level, alarms, serial numbers, internal battery level and User ID (by customer request)
• System failure
• Probe failure (probe disconnected)
• Low internal battery
• High temperature (38.0° C or 40.0° C)
• Sensor failure
• Skin over temperature (37.5° C or 38.5° C)
• Mains failure
• Air over temperature (37.5° C or 39.5° C)
• Blower failure
• Warm-up failure (> 30 minutes)
• Airflow failure
• Skin deviation ±1°C and air deviation ±3°C
• Portable phototherapy ( TLC 2 HP)
• Oxyhood (TSN 046 series)
• Digital baby scale ( TSN 9806 DBS)
• X-ray cassette tray
• Double wall